5
Introduction to the Nano Pump
The Main Power Supply Assembly
The main power supply comprises a closed assembly (no onsite repair
possibility).
The power supply provides all DC voltages used in the module except for the
voltages supplied by the lamp power supply to the deuterium and tungsten
lamps in the detectors. The line voltage can vary in a range from 100 240 volts
AC ± 10 % and needs no manual setting.
Line
voltage
Figure 49
To disconnect the instrument from line, unplug the power cord. The power
supply still uses some power, even if the power switch on the front panel is
turned off.
228
MPS
primary
circuits
sense
(*) total power consumption on +36 V and +24 V must not exceed 107 watts.
Main Power Supply (MPS) Block diagram
+12 V
supply
floating
+24 V
supply
+5 V
switching
regulator
linear
voltage
regulators
1100 Series Nano Pump Service Manual
+36 V / 2.5 A (*)
+24 V / 4.5 A (*)
+5 V / 3 A
+15 V /.3 A
-15 V /.3 A