To Install A Capillary Column In A Split/Splitless Inlet - Agilent Technologies MSD 5977B Series Operating Manual

Table of Contents

Advertisement

2
Installing 8890 GC Columns

To Install a Capillary Column in a Split/Splitless Inlet

To Install a Capillary Column in a
Split/Splitless Inlet
Materials needed
To install columns in other types of inlets, refer to your GC User Information.
The GC operates at high temperatures. Do not touch any parts of the GC until
WAR N IN G
you are sure they are cool.
Always wear safety glasses when handling capillary columns. Avoid
WAR N IN G
puncturing your skin with the end of the column.
Always wear clean gloves while handling any parts that go inside the GC or
C AU T I O N
analyzer chambers.
38
Gloves, clean
Large (8650-0030)
Small (8650-0029)
Metric ruler
Wrench, open-end, 1/4-inch and 5/16-inch (8710-0510)
Capillary column
Column cutter, ceramic (5181-8836) or diamond (5183-4620)
Ferrules
0.27-mm id, for 0.10-mm id columns (5062-3518)
0.37-mm id, for 0.20-mm id columns (5062-3516)
0.40-mm id, for 0.25-mm id columns (5181-3323)
0.5-mm id, for 0.32-mm id columns (5062-3514)
0.8-mm id, for 0.53-mm id columns (5062-3512)
Inlet column nut (5181-8830 for Agilent 7890/7820 and 8890/8860 GCs)
Magnifying loupe
Septum (may be old, used inlet septum)
5977B Series MSD Operating Manual