2
Installing 8890 GC Columns
To Install a Capillary Column in a Split/Splitless Inlet
To Install a Capillary Column in a
Split/Splitless Inlet
Materials needed
•
•
•
•
•
•
•
•
•
To install columns in other types of inlets, refer to your GC User Information.
The GC operates at high temperatures. Do not touch any parts of the GC until
WAR N IN G
you are sure they are cool.
Always wear safety glasses when handling capillary columns. Avoid
WAR N IN G
puncturing your skin with the end of the column.
Always wear clean gloves while handling any parts that go inside the GC or
C AU T I O N
analyzer chambers.
38
Gloves, clean
•
Large (8650-0030)
•
Small (8650-0029)
Metric ruler
Wrench, open-end, 1/4-inch and 5/16-inch (8710-0510)
Capillary column
Column cutter, ceramic (5181-8836) or diamond (5183-4620)
Ferrules
•
0.27-mm id, for 0.10-mm id columns (5062-3518)
•
0.37-mm id, for 0.20-mm id columns (5062-3516)
•
0.40-mm id, for 0.25-mm id columns (5181-3323)
•
0.5-mm id, for 0.32-mm id columns (5062-3514)
•
0.8-mm id, for 0.53-mm id columns (5062-3512)
Inlet column nut (5181-8830 for Agilent 7890/7820 and 8890/8860 GCs)
Magnifying loupe
Septum (may be old, used inlet septum)
5977B Series MSD Operating Manual