2
Troubleshooting and Test Functions
.
8 The test results are evaluated automatically.
Test results
In the ChemStation, the measured values are evaluated automatically. A table
with the measured pressure drop over the inlet filter, the measured flow and
pressure offset and the correction factors for all steps are printed.
You have now the possibility to: accept and store all the correction factors and
the measured offset, accept only the measured offset, reset (no correction
factor or offset are applied).
88
1100 Series Nano Pump Service Manual