Test evaluation
Figure 11
Table 5
Flow sensor accuracy calibration test evaluation
Pressure drop over filter
Offset flow
Offset pressure sensor
1100 Series Nano Pump Service Manual
Flow sensor accuracy calibration test evaluation
The measured value should not be higher
than 10 Bar
This value can be positive or negative
and should be between 500 nL/min and
-90 nL/min.
This value has not to be considered in the
evaluation.
Troubleshooting and Test Functions
If higher we recommend to
change the filter in front of the
EMPV and restart the test again.
*
If higher or lower replace the
nano flow sensor.
2