To fill the pump with oil, unscrew the black cap to the right of the exhaust
filter. Using the oil (1990-924-00) and the funnel (1331-191-00) provided
in the installation kit, slowly fill the pump until the level in the sight glass
is approximately 1/4" below the fill line. Replace the black oil fill cap.
When all shipping material has been removed, all pump connections have
been made and the pump has been filled with oil, replace the right side
panel and close the front and rear doors.
An Installation Kit (VIC P/N 0137-800-00) is included with the unit when it
is shipped. The kit contains:
VIC Leak Detection
2099 Ninth Avenue
Ronkonkoma, NY 11779
631.737.0900
1109-186-00 Rev. C
NOTE:
The MS-40 Leak Detector is factory tested and calibrated
prior to shipment and once installed, is ready for use.
Part Number
1680-055-00
0103-141-01
1331-191-00
1330-391-00
1990-924-00
0137-807-00
0137-804-00
1621-224-02
1621-227-05
0126-284-00
1620-731-00
1750-054-00
Description
Exhaust Filter
Filament Kit (box of 5)
Funnel
7/64 Allen Wrench
VMP-19 Oil (1 liter)
Kit, Cart Mounting
N2 Kit
Adapter, KF16 to 1/4" NPT
QC Adapter, KF25 to 1 1/8"
Blank Plug, 1/18
Elbow, 1/4 NPT to 3/8 Tube
Tubing, Tygon, 1/4"
1-5