Veeco MS40 Operation And Maintenance Manual page 7

Portable automatic leak detector
Table of Contents

Advertisement

the back panel must be opened. Both panels are hinged at the bottom and
held with a single screw at the top of each panel. To open a panel, loosen
the captive screw at the top of the panel (a 1/4 turn is sufficient to unloosen
the screw) and gently lower the door. Make sure that the surface the unit is
placed on a flat surface (VIC Cart, P/N 0137-705-00, recommended).
Remove all shipping materials and tie-down wraps from the internal
pumps.
The MS-40 is shipped without oil in the internal roughing pump, which must
be filled prior to use. In addition, to protect the pump during shipment, it is
fitted with two blanks that cover its intake and filter connections. Remove
these blanks and attach the flex hosing and filter (refer to Figure 1-2 for a
diagram illustrating the proper installation for the hose and filter connections)
with the provided centering rings and clamps. Also, the unit's purge tubing
(refer to Figure 1-2) is fitted with a 1/4" ferrule blank. Unscrew the blank and
reattach the tubing as indicated in Figure 1-2.
Parts List for Figure 1-2:
ILLUSTRATION # PART NUMBER
Vacuum Instrument Corp.
2099 Ninth Avenue
Ronkonkoma, NY 11779
631.737.0900
1109-186-00 Rev. C
1
1660-362-00 110/120v
1660-362-01 220/240v
2
1621-183-00
3
1750-046-00
4
1620-726-00
5
1621-231-12
6
1621-239-02
7
0137-205-00
8
1680-055-00
9
1620-743-00
DESCRIPTION
(60Hz)
(50 Hz) Pump
Ferrule Nut
Poly-flo Tubing, Purge Line
Elbow / with ferrule nut
KF16 Centering Ring
KF16 Clamp
Manifold, Roughing Pump
Exhaust Filter
KF16 to 1/4 NPT Adapter
3/8 MPT x ¼ Bushing
1-4