'U' - Followed by 4 to 5 characters.
'V' - Followed by 1 to 4 numbers.
'W' - Followed by 4 or 5 characters.
'X' - Followed by 4 characters.
Vacuum Instrument Corp.
2099 Ninth Avenue
Ronkonkoma, NY 11779
631.737.0900
1109-186-00 Rev. C
Direct Flow Gain (G3).
(e.g. 2.00 = 2.00 x gain)
(e.g. 10.00 = 10.00 x gain)
In software revisions 3.8 and later, this field is fixed at 5 characters.
('U' + 4 numbers).
(e.g. 0971 = 9.71 x gain)
Crossover #1 or Reverse Flow Crossover in millitorr.
(e.g. 10 = 10 millitorr)
(e.g. 200 = 200 millitorr)
(e.g. 9000 = 9000 millitorr)
In software revisions 3.8 and later, this field is fixed at 5 characters.
('V' + 4 numbers).
(e.g. 0300 = 300 millitorr)
Stamped Leak Rate on Internal Calibrator in atm-cc/sec when state of
leak detector is in any state except I-47. When the leak detector is
in state I-47, this field will be the temperature compensated value of the
internal calibrator. This value is what the leak detector is calibrating to
when the internal calibrator is selected as the active calibrator.
(e.g. 5708 = 5.7 x 10-8 atm-cc/sec)
(e.g. 10010 = 10.0 x 10-10 atm-cc/sec)
In software revisions 3.8 and later, this field is fixed at 5 characters.
('W' + 4 numbers).
(e.g. 6304 = 6.3 x 10-4 atm-cc/sec. A mantissa of 10.0 will never be
sent.)
Stamped Temperature on Internal Calibrator in Degrees Celsius.
(e.g. 20.2 = 20.2 °C)
(e.g. 35.4 = 35.4 °C)
In software revisions 3.8 and later, this field is fixed at 3
('X' + 2 numbers).
(e.g. 25 = 25°C)
d-viii
characters.