Veeco MS40 Operation And Maintenance Manual page 150

Portable automatic leak detector
Table of Contents

Advertisement

'd' - Followed by 4 numbers.
'e' - Followed by 3 numbers.
'f' - Followed by 3 numbers.
'g' -
All revisions prior to Rev 4.0 transmit the serial stream as follows:
Vacuum Instrument Corp.
Vacuum Instrument Corp.
2099 Ninth Avenue
2099 Ninth Avenue
Ronkonkoma, NY 11779
Ronkonkoma, NY 11779
631.737.0900
631.737.0900
1109-186-00 Rev. C
Reverse Rough Close Delay (D7) in Tenths of seconds.
(e.g. 0001 = 0.1 seconds)
(e.g. 0142 = 14.2 seconds)
This field did not exist In the serial stream in software revisions
prior to
Rev. 4.1. However, is was present on the MS-40 user panel and
functioned properly in software Revisions 3.9 and later.
Profile Switch Setting (0 = 255)
(e.g. 001)
(e.g. 128)
This field did not exist In the serial stream in software
revisions prior to
Rev. 4.1.
Revision Switch Setting (0 = 255)
(e.g. 001)
(e.g. 128)
This field did not exist In the serial stream in software
revisions prior to
Rev. 4.1.
Followed by 6 numbers. (first 5 are the leak rate and the last is reject
status). Previous Test Final Leak Rate Value & Reject Status measured
in atm-cc/sec or mbar-l/sec. See 'H' for reject status values. This is the
leak rate value and reject status of the previous leak test cycle just prior
to the termination of the test. It is used for data logging.
(e.g. 057082 = 5.7 x 10-8 atm-cc/sec & Accept or Pass)
(e.g. 100103 = 10.0 x 10-10 atm-cc/sec & Reject or Fail)
This field did not exist In the serial stream in software revisions prior to
Rev. 4.2.
Packets 'C' through 'I' are sent all the time. Packets 'A' and 'B' are
only sent when necessary. Packets 'L' and 'N' are send every third
time after packet 'I'. Packets 'J', 'K', and 'O' through 'g' are sent
every 33 times after packets 'L' and 'N'.