Starting The Ms-40 Leak Detector - Veeco MS40 Operation And Maintenance Manual

Portable automatic leak detector
Table of Contents

Advertisement

2.3
The MS-40 is started from the User Panel simply by pressing the ON/OFF
(I/O) switch. During the three-minute Start-up mode, the following will
occur within the unit:
NOTE:
During the start up procedure, all LED's on the User Panel will be tested
by being sequentially lit.
When the Start-up mode has been completed, the Automatic Ranging
indicator will be lit, the test port measurement gauge will be activated, the
User Panel numeric display will indicate 4 and the READY light
indicator will be lit.
Pre-Testing Start-Up Procedure:
Ensure that all connections are secure and all utilities are available. Check
that:
Vacuum Instrument Corp.
2099 Ninth Avenue
Ronkonkoma, NY 11779
631.737.0900
1109-186-00 Rev. C

Starting the MS-40 Leak Detector

1
The unit will activate a self-diagnostic routine.
2
The fore-line area will be pumped down.
3
The turbo-molecular pump will spin up to its rated speed.
4
If the filament was on when the unit was last turned off, the
filament will be on again at the end of the start-up mode. If the
filament was off when the unit was turned off, the user will
have to press the FIL ON momentary switch to turn on the
filament.
A.C. power is connected.
A helium supply source is ready and available (if
applicable).
There is a sufficient supply of oil in both the internal and
external rough pumps
All external accessories are properly connected.
2-19

Hide quick links:

Advertisement

Table of Contents
loading

Related Products for Veeco MS40

Table of Contents