2-14
Calibration
Precalibration
1.
2.
Figure 2-3
Connections for factory DC volts calibration
Model 2182
NOTE
3.
Table 2-7
Remote precalibration steps
Precalibration step
10V
0V
1V
NOTE
Connect the calibrator to CHANNEL 1 of the Model 2182 input jack using the low-
thermal cable, as shown in
before proceeding. (See
Table 2-3
Set the calibrator to output DC volts, and turn external sense off.
2107 Low
Red (CH1)
Thermal Cable
Green (CH2)
CHANNEL 1
!
CHANNEL 2
Black (CH1)
White (CH2)
Keep drafts away from low-thermal connections to avoid thermal drift, which could
affect calibration accuracy.
Perform the precalibration steps summarized in
•
Set the calibrator to the indicated voltage, and make sure the unit is in operate.
•
Allow the signal voltage to settle.
•
Send the indicated programming command.
•
Wait until the Model 2182 completes each step before continuing.
Calibrator voltage
10.000000V
0.0000000V
1.0000000V
Keep the calibrator locked to the 11V range.
Figure
2-3. Wait two minutes to allow for thermal equilibrium
for Model 2107 cable connections.)
OUTPUT HI
DC Voltage Calibrator
OUTPUT LO
Table
Calibration command
:CAL:PROT:PCAL:STEP0 10
:CAL:PROT:PCAL:STEP1 0
:CAL:PROT:PCAL:STEP2 1
2-7. For each step: