Analog Output; Remote Interface - Keithley 2182 Service Manual

Nanovoltmeter
Hide thumbs Also See for 2182:
Table of Contents

Advertisement

2182A Nanovoltmeter Specifications

ANALOG OUTPUT

MAXIMUM OUTPUT: ±1.2V.
ACCURACY: ±(0.1% of output + 1mV).
OUTPUT RESISTANCE: 1kΩ ±5%.
-9
6
GAIN: Adjustable from 10
to 10
. With gain set to 1, a full range input will produce a
1V output.
OUTPUT REL: Selects the value of input that represents 0V at output. The reference value
can be either programmed value or the value of the previous input.
TRIGGERING AND MEMORY
WINDOW FILTER SENSITIVITY: 0.01%, 0.1%, 1%, 10%, or full scale range (none).
READING HOLD SENSITIVITY: 0.01%, 0.1%, 1%, or 10% of reading.
TRIGGER DELAY: 0 to 99 hours (1ms step size).
EXTERNAL TRIGGER DELAY: 2ms + <1ms jitter with auto zero off, trigger delay = 0.
MEMORY SIZE: 1024 readings.
MATH FUNCTIONS
Rel, Min/Max/Average/Std. Dev/Peak-to-Peak (of stored reading), Limit Test, %, and mX+b
with user defined units displayed.

REMOTE INTERFACE

Keithley 182 emulation.
GPIB (IEEE-488.2) and RS-232C.
SCPI (Standard Commands for Programmable Instruments).
Notes:
1. Relative to calibration accuracy.
2. With Analog Filter on, add 20ppm of reading to listed specification.
3. When properly zeroed using REL function. If REL is not used, add 100nV to the range
accuracy.
4. Specifications include the use of ACAL function. If ACAL is not used, add 9ppm of
reading/°C from T
to the listed specification. T
CAL
during ACAL.
5. For 5PLC with 2-reading Digital Filter. Use ±(4ppm of reading + 2ppm of range) for
1PLC with 10-reading Digital Filter.
6. Channel 2 must be referenced to Channel 1. Channel 2 HI must not exceed 125%
(referenced to Channel 1 LO) of Channel 2 range selected.
7. Noise behavior using 2188 Low Thermal Short after 2.5 hour warm-up. ±1°C. Analog
Filter off. Observation time = 10X response time or 2 minutes, whichever is less.
On, line frequency ±0.1%. If L
8. For L
SYNC
9.
For 1kΩ unbalance in LO lead. AC CMRR is 70dB.
10. For Low Q mode On, add the following to DC noise and range accuracy at stated
response time: 200nV p-p @ 25s, 500nV p-p @ 4.0s, 1.2µV p-p @ 1s, and 5µV p-p
@ 85ms.
is the internal temperature stored
CAL
Off, use 60dB.
SYNC
GENERAL SPECIFICATIONS
POWER SUPPLY: 100V/120V/220V/240V.
LINE FREQUENCY: 50Hz, 60Hz, and 400Hz, automatically sensed at power-up.
POWER CONSUMPTION: 22VA
OPERATING ENVIRONMENT: Specified for 0° to 50°C. Specified to 80% RH at 35°C.
MAGNETIC FIELD DENSITY: 10mV range 4.0s response noise tested to 500 gauss.
STORAGE ENVIRONMENT: -40° to 70°C.
WARRANTY: 3 years.
SAFETY: Complies with European Union Directive 73/23/EEC (low voltage directive);
meets EN61010-1 safety standard. Installation category I.
EMC: Complies with European Union Directive 89/336/EEC (CE marking requirement),
FCC part 15 class B, CISPR 11, IEC 801-2, IEC-801-3, IEC 801-4.
VIBRATION: MIL-PRF-28800E Type III, Class 5.
WARM-UP: 2.5 hours to rated accuracy.
DIMENSIONS: Rack Mounting: 89mm high x 213mm wide x 370mm deep (3.5in x
8.375in x 14.563in). Bench Configuration (with handles and feet): 104mm high x 238mm
wide x 370mm deep (4.125 in x 9.375 in x 14.563 in).
SHIPPING WEIGHT: 5kg (11 lbs).
ACCESSORIES SUPPLIED
2107-4: Low Thermal Input Cable with spade lugs, 1.2m (4 ft).
User manual, service manual, contact cleaner, line cord, alligator clips.
ACCESSORIES AVAILABLE
2107-30:
Low Thermal Input Cable with spade lugs, 9.1m (30 ft)
2182-KIT:
Low Thermal Connector with strain relief
2188:
Low Thermal Calibration Shorting Plug
2187-4
Input Cable with safety banana plugs
4288-1:
Single Fixed Rack Mount Kit
4288-2:
Dual Fixed Rack Mount Kit
7007-1:
Shielded GPIB Cable, 1m (3.2 ft)
7007-2:
Shielded GPIB Cable, 2m (6.5 ft)
7009-5:
Shielded RS-232 Cable 1.5m (5 ft)
8501-1:
Trigger-Link Cable 1m (3.2 ft)
8501-2:
Trigger-Link Cable 2m (6.5 ft)
8502:
Trigger-Link Adapter to 6 female BNC connectors
8503:
Trigger-Link Cable to 2 male BNC connectors
11. After 2.5 hour warm-up, ±1°C, 5PLC, 2 minute observation time, Channel 1 10mV
range only.
12. For Channel 1 or Channel 2, add 0.3°C for external reference junction. Add 2°C for
internal reference junction.
13. Speeds are for 60Hz (50Hz) operation using factory defaults operating conditions
(*RST). Autorange Off, Display Off, Trigger Delay = 0, Analog Output off.
14. Speeds include measurements and binary data transfer out the GPIB. Analog Filter On,
4 readings/s max.
15. Auto Zero Off, NPLC = 0.01.
16. 10mV range, 80 readings/s max.
17. Sample count = 1024, Auto Zero Off.
18. For L
On, reduce reading rate by 15%.
SYNC
19. For Channel 2 Low Q mode Off, reduce reading rate by 30%.
20. Front Auto Zero Off, Auto Zero Off.
21. Applies to measurements of room temperature resistances <10Ω, Isource range ≤ 20uA
22. Display off, delay 1ms
RKN 6/08/04
Rev A
Page 3 of 3

Advertisement

Table of Contents
loading

This manual is also suitable for:

2182a

Table of Contents