Saving calibration constants
After completing the calibration procedure, send the following command to save the new
calibration constants:
:CAL:PROT:SAVE
NOTE
Calibration constants will not be saved unless the :SAVE command is sent.
Calibration constants can be returned using the :CAL:PROT:DATA? command. For
details, see
"Miscellaneous
Default calibration constants and tolerances are listed in
Locking out calibration
After saving calibration, send the following command to lock out calibration:
:CAL:PROT:LOCK
commands,
:DATA?"
in
Appendix
Calibration
2-17
B.
Appendix B
(see
Table
B-2).