Keithley 2182 Service Manual page 41

Nanovoltmeter
Hide thumbs Also See for 2182:
Table of Contents

Advertisement

2-16
Calibration
Analog output calibration
1.
2.
Figure 2-5
Connections for analog output calibration
SHIFT
LOCAL
POWER
Programming calibration dates
Program the present calibration date and calibration due date by sending the following
commands:
:CAL:PROT:DATE <year>, <month>, <day>
:CAL:PROT:NDUE <year>, <month>, <day>
For example, the following commands assume calibration dates of 12/15/97 and 12/15/98
respectively:
:CAL:PROT:DATE 1997, 12, 15
:CAL:PROT:NDUE 1998, 12, 15
Connect the Model 2182 analog output to CHANNEL 1 of the front panel input jack, as
shown in
Figure
2-5.
Send the following command to calibrate the analog output:
:CAL:PROT:DC:STEP5
Connect to Clip Leads
ANALOG OUTPUT
Jack on Rear Panel
REM
STEP
SCAN
CH1
CH2
CH3
CH4
CH5
CH6
CH7
CH8
CH9
CH10
MATH
TALK
REAR
LSTN
SRQ
SHIFT
4W
TIMER
HOLD
TRIG
FAST
MED
SLOW
REL
FILT
AUTO
ERR
BUFFER
STAT
2182 NANOVOLTMETER
%
L
MX+B
V
1
-V
2
SYNC
TYPE
OUTPUT
A
TCOUPL
OUT
/
TEMP
TEMP
DCV1
DCV2
V
1
V
2
ACAL
FILT
REL
1
2
HOLD
BUFFER
LIMITS
DELAY
EX TRIG
TRIG
STORE
VALUE
ON/OFF
RECALL
CONFIG
HALT
SETUP
GPIB
RS232
CAL
TEST
STEP
SCAN
SAVE
RESTR
DIGITS RATE
EXIT
ENTER
MODEL 2182
HI
LO
CHANNEL 1
LO
HI
!
HI
LO
CHANNEL 2
120V MAX
RANGE
AUTO
12V MAX
RANGE
CAT I
350V PEAK ANY
TERMINAL TO CHASSIS
Connect
2107 Low
Thermal Cable
Spade Lugs
to Clip Leads

Advertisement

Table of Contents
loading

This manual is also suitable for:

2182a

Table of Contents