Keithley 2182 Service Manual page 36

Nanovoltmeter
Hide thumbs Also See for 2182:
Table of Contents

Advertisement

NOTE
2.
3.
DC volts calibration
After front panel short and open steps, do the following:
1.
2.
Table 2-5
DC voltage remote calibration programming steps
Calibration step Calibrator voltage
+10V
+10.000000V
-10V
-10.000000V
* Use indicated values whenever possible. Change parameter accordingly if using a different calibrator voltage.
Programming calibration dates
Program the present calibration date and calibration due date by sending the following
commands:
:CAL:PROT:DATE <year>, <month>, <day>
:CAL:PROT:NDUE <year>, <month>, <day>
For example, the following commands assume calibration dates of 12/15/97 and 12/15/98
respectively:
:CAL:PROT:DATE 1997, 12, 15
:CAL:PROT:NDUE 1997, 12, 15
Keep drafts away from low-thermal connections to avoid thermal drift, which could
affect calibration accuracy.
Send the following command:
:CAL:PROT:DC:STEP1
After the Model 2182 completes this step, remove the low-thermal short, and send this
command:
:CAL:PROT:DC:STEP2
Connect the calibrator to the CHANNEL 1 input of the Model 2182 as shown in
Figure
2-2.
(SeeTable 2-3
Perform the calibration steps summarized in
Set the calibrator to the indicated voltage, and make sure the unit is in operate. (Use
the recommended voltage whenever possible.)
Allow the signal voltage to settle.
Send the indicated programming command. (Change the voltage parameter if you
are using a different calibration voltage.)
Wait until the Model 2182 completes each step before continuing.
Calibration command*
:CAL:PROT:DC:STEP3 10
:CAL:PROT:DC:STEP4 –10
for wire colors.) Allow two minutes for thermal equilibrium.
Table
2-5. For each step:
Calibration
2-11
Parameter range
9 to 11
-9 to –11

Advertisement

Table of Contents
loading

This manual is also suitable for:

2182a

Table of Contents