B-8
Calibration Reference

:DATE

(:CALibration:PROTected:DATE)
Purpose
Format
Parameters
Query
Response
Description
NOTE
Example

:NDUE

(:CALibration:PROTected:NDUE)
Purpose
Format
Parameters
Query
Response
Description
NOTE
Example
To program the calibration date.
:cal:prot:date <year>, <month>, <day>
<year> = 1997 to 2096
<month> = 1 to 12
<day> = 1 to 31
:cal:prot:date?
<year>, <month>, <day>
The
command allows you to store the calibration date in instrument
:DATE
EEROM for future reference. You can read back the date from the instrument
by using the
:DATE?
The year, month, and day parameters must be delimited by commas.
:CAL:PROT:DATE 1997,11,20
To send the next calibration due date to the instrument.
:cal:prot:ndue <year>, <month>, <day>
<year> = 1997 to 2096
<month> = 1 to 12
<day> = 1 to 31
:cal:prot:ndue?
<year>, <month>, <day>
The
command allows you to store the date when calibration is next
:NDUE
due in instrument memory. You can read back the next due date by using the
query.
:NDUE?
The next due date parameters must be delimited by commas.
:CAL:PROT:NDUE 1998,11,20
query.
Send cal date (11/20/97).
Send due date (11/20/98).