Calibration / Configuration Via Display / Keypad - Ametek Drexelbrook Universal IV Quick Start Manual

Cut monitor model transmitters 2-wire rf admittance / capacitance level measurement system with hart protocol
Table of Contents

Advertisement

2.5

Calibration / Configuration Via Display / Keypad

Menu Function
(display
abbreviation)
"Fct 1.00
Water Cut Ranges
(RANGE)"
"Fct 2.00
Calibration (CAL)"
"Fct 2.01
Indicated Calibration Point
(IND CAL)"
"Fct 2.02
Actual Calibration Point
(ACT CAL)"
"Fct 3.00
Strapping table
(STRAP)"
"Fct 3.01
Maximum points
(MAX PNT)"
"Fct 3.02
Point number index
(INDEX)"
• All function in the PC software except Save/Print re
accessible via the display/keypad
• To enter the Configuration Menu follow these steps
• Press and hold the ENTER button for approximately 5
seconds
• Use the UP and DOWN buttons to scroll through the
available selections
• Press ENTER to access sub-menu items
• Use UP and DOWN buttons to adjust settings - settings
that can be adjusted will be flashing
• Press ENTER to accept the adjustment or press and
hold ENTER for 5 seconds to exit to previous menu
"Values
(display
abbreviation)"
0 to 1% water in Light Oil (LIGHT A)
0 to 5% water in Light Oil (LIGHT B)
0 to 10% water in Light Oil (LIGHT C)
0 to 30% water in Light Oil (LIGHT D)
0 to 50% water in Light Oil (LIGHT E)
0 to 1% water in Heavy Oil (HEAVY F)
0 to 5% water in Heavy Oil (HEAVY G)
0 to 10% water in Heavy Oil (HEAVY H)
0 to 30% water in Heavy Oil (HEAVY I)
0 to 50% water in Heavy Oil (HEAVY J)
0 to 80% water in Heavy Oil (HEAVY M) -
Default
CUSTOM
% water - 0.0 Default
% water - 0.0 Default
15 (Default)
"1..MAX PNT
1 (Default)"
Select the water cut range for optimal
measurement
'Light Oil' is defined as oil with API Gravity
less than 25. Heavy Oil is defined as oil with
API Gravity greater than 25.
Custom range requires a custom strapping
table. See Fct 3.00
Enter this menu to calibrate the unit
Enter the water cut reading captured at the
time of taking the sample for calibration
measurement
Enter the actual water cut reading verified by
another method of water cut measurement
Use this strapping table menu to define
a custom range if selected in FCT 1.00.
Otherwise the correct strapping table is
automatically loaded when range is selected in
FCT 1.00.
Default values are for 'HEAVY M'
range
Enter the total number of points in the
strapping table which is range dependent
Enter the point number index
15
Quick Start Guide
Description

Advertisement

Table of Contents
loading

Table of Contents