Ametek Drexelbrook Universal IV Quick Start Manual page 21

Cut monitor model transmitters 2-wire rf admittance / capacitance level measurement system with hart protocol
Table of Contents

Advertisement

2.5
Calibration / Configuration via Display / Keypad (Continued)
"Fct 6.01
Restore factory default
(RST FAC)"
"Fct 6.02
Pad Capcitor in PF
(PAD CAP)"
"Fct 6.03
Contrast
(CONTRST)"
"Fct 6.04
Parameter Number
(PAR NUM)"
"Fct 6.05
Parameter Offset
(PAR OFS)"
"Fct 6.06
Parameter Value
(PAR VAL)"
NO (Default)
30.0 (Default)
0 (Default)
0
0
44
Select YES to restore factory default in which
case all paramaters will be replaced with
factory default setting. Restoring the factory
default will initiate this message on the display
'DEFAULT PARAMS SET' until power is cycled.
Enter the value of an external capacitor
that must be connected to the unit. Padding
capacitors are used to reduce the sensing
element standing capacitance in order to
improve the measurement resolution
0 is the highest contrast. 20 is the lowest
contrast
0 to 65535. Contact factory
Contact factory
Contact factory
17
Quick Start Guide