Section 3: Configuring the Device
3.0 Calibration & Configuration
3.1 Calibration of the Cut Monitor Module
Normally the Cut Monitor transmitter will not require calibration. All that is required is it
be configured with the correct Water Cut Range.

3.2 Operating the User Interface

Configuration and calibration of the DCM is performed via the display keypad or STExplorerTM
PC software. Reference "Menu Tree" (Ref 3.6.) and "Configuration Menu Table" (Ref 3.5).
- When a button is pressed a visual indication appears on the left of the display (Bar).
- To access the menu, press and hold the "ENTER" button for approximately 5 seconds. Note
visual indication on left side display increments at 1 bar/second.
- Use the "UP" and "DOWN" buttons to scroll through the available menu selection items.
- Press "ENTER" to access sub-menu items
- To "EDIT" a value, press the "ENTER" button once – Value will flash on last digit
- Navigate left (Back) or right (Enter) to highlight the appropriate character
- Settings that can be changed will be flashing
- In numeric fields pressing "ENTER" will move the flashing digit to the right. When the far
right digit is flashing pressing "ENTER" accepts the entry
- Press "ENTER" to accept the change.
- In numeric fields pressing back will move the flashing digit to the left.
- Press the "BACK" button to exit to the previous menu level
- Pressing and holding the "BACK" button for 5 seconds exits the menu to the display mode.
Note visual indication on left side display increments at 1 bar/second.
- The unit will automatically exit the menu and return to normal display if no input is received
in 60 approximately seconds
24