Manual Phase Measurement - AEMC OX7102-CII User Manual

Portable oscilloscope ox-cii series
Table of Contents

Advertisement

Applications
Applications (cont'd)
6. Manual measurement
of phase
Reminder
7. Display of a TV video
signal
SCOPIX Oscilloscopes
• Select manual phase measurement via the menu:
Measure
Manual phase measurement (see §. Measure Menu).
∗ The 2 cursors (1 and 2) of the manual measurements are displayed on
the reference trace. A "ϕ" cursor in relation to which the phase
is measured is displayed.
∗ The phase measurement (in °) is indicated under the display of the
curve(s).
Example: (1)Ph = 150.0°
The 3 measurement cursors are present if at least one trace is present
on the screen.
The 3 measurement cursors can be moved directly on the screen
by using the stylus. They can also be moved using the stylus by
selecting 1 (cursor 1) or 2 (cursor 2) in the bargraph in the
status zone.
If the unattached cursors option is not activated (see §. Unattached
cursors), the cursors (1 and 2) will remain linked to the trace when moved.
If the option is active, these cursors can be moved as required on the
screen.
In all cases, the symbol "ϕ" can be moved freely.
This example illustrates the TV synchronization functions.
- The parameters in the TV menu (Trigger Parameters menu) for
displaying a TV signal only apply to the CH1 input.
- it is recommended to use a 75
• Use a Probix HX0031 adapter to inject on CH1 a composite TV signal with
the following characteristics:
- 625 lines
- positive modulation
- vertical grey scale stripes
• In the Trigger Parameters menu, select the tab:
Trig. Menu
Parameter
• Set the standard number of lines: 625 lines
polarity: +
-
line: 25 (for a video signal)
-
-
front: + (or using the key opposite).
• Adjust the CH1 coupling:
Vert Menu
CH1
or by pressing the AC/DC GND key.
• Adjust the CH1 sensitivity:
Vert Menu
CH1
or using the keys opposite.
• Adjust the scan speed:
using the scrollbar in the time base window: 20 µs/div.
or using the keys opposite.
Ω
adapter for observing a video signal.
TV:
Sensitivity/coupling
Sensitivity/coupling
DC
200 Mv/div.
IX - 161