OtoRead™ Instructions for Use - US
Reference equivalent threshold values for transducer
Output
Frequency (Hz)
Frequency (Hz)
732.4
1037.6
1464.8
2075.2
2929.7
4150.4
5859.4
Pin assignments
The probe connector pin out:
Pin 1
Rcvr +
Pin 2
Rcvr Shield
Pin 3
Rcvr -
Pin 4
Reserved
Pin 5
Shield
Pin 6
Reserved
Pin 7
Mic Power +
Pin 8
Mic Shield
Pin 9
Mic Out
Pin 10
Mic Power -
Table 2: Frequency and Intensity with G.R.A.S. RA0045 OES
Minimum
Maximum
Frequency (Hz)
727
737
1033
1043
1460
1470
2070
2080
2925
2935
4145
4155
5855
5865
Table 3: Probe Nominal Sound Channel Magnitudes in dB SPL
Frequency [Hz]
732.4
1037.6
1464.8
2075.2
2929.7
4150.4
5859.4
Type A receptacle HDMI (female)
Minimum
Magnitude (dB SPL)
83
85
88
92
92
85
76
IEC 60711, RA-0045
88.0
90.0
93.5
97.8
97.8
90.6
81.9
Pin 11
Unused
Pin 12
Unused
Pin 13
Unused
Pin 14
Reserved
Pin 15
Comm Power
Pin 16
Comm Data
Pin 17
Ground
Pin 18
+3.3V
Pin 19
Ground
Page 49
Maximum
Magnitude (dB SPL)
93
95
98
102
102
95
86

Advertisement

Table of Contents
loading

Table of Contents