3.11.4.5 Minimum value

3.11.4.6 Clock mode

3.11.4.7 Graph style

The Minimum (Amplitude) Value setting allows the user to set the unit
to include minimum amplitude values in the pass/refer criterion. The
TM
OtoRead
is set with this feature turned OFF when it is shipped from
the factory. If the MIN VALUE is set to ON, a result is not considered a
pass unless the amplitude at each frequency is equal to or greater
than the minimum value programmed into the unit. This is in addition
to meeting the other pass criteria, including the minimum SNR and the
number of passing frequencies for overall test Pass.
To change the mode to Minimum Amplitude setting, press the LEFT
or RIGHT arrow keys to make a selection. Once you have made your
selection, press
NEXT.
The minimum DP amplitude when ON is -5 dB SPL.
The minimum TE amplitude selections are -5 dB SPL. and -10 dB
SPL.
The Clock Mode menu allows the user to change the clock from a 24
hour mode to a 12 hour mode. To change the clock mode, press the
CHANGE keys. Press the
The Graph Style menu allows the user to select from two options for
viewing the results. The SNR graph view shows the signal-to-ratio for
each DP test frequency or TE test band. The Value graph view shows
the absolute emission and noise levels for each DP test frequency or
TE test band.
NEXT to exit this menu.
Page 34