Troubleshooting - Interacoustics OtoRead Instructions For Use Manual

Hide thumbs Also See for OtoRead:
Table of Contents

Advertisement

OtoRead™ Instructions for Use - US

5 Troubleshooting

Problem
Instrument does not
turn on
The test will not start
The results will not
print
Display is frozen and
instrument will not
respond to button
presses
Attach
Probe
BT Device
Not Found
BT Error
#xxx
BT Not
Configured
Device Not
Responding
Solutions
The
DOWN arrow must be pressed for a full second (the Yellow
'TEST' LED will illuminate)
Connect the charger as shown in chapter 2.6.2. Confirm that the blue
'Charging' LED is illuminating in a slow blink pattern. Wait at least 10
minutes and then attempt to turn on the instrument
Contact Interacoustics or your local distributor if the problem persists
Select a different sized ear tip
Reposition the probe
Change the probe tube
Verify that the ear tip is sealed in the ear canal via feedback from the
PROBE CHECK screen
Check that the instrument will start in your own ear with the proper ear
tip for testing yourself. If the test will not start or if the AutoStart tones
sound unusual, replace the probe tube
Contact Interacoustics or your local distributor for service if the problem
persists across several patients
Check the printer status. Turn the printer on (wake from sleep mode) by
pressing the large button
If the printer does not turn on, plug in the power supply to charge the
battery
Be sure the printer has paper
If paper comes out of the printer but there is no text on the paper then
the paper is in backwards
Press the large printer button twice, rapidly, to run demo print
Contact Interacoustics or your local distributor for service if the problem
persists
Press and hold the
instrument to power off. Powering the instrument back on again should
reset/restore normal function
Contact Interacoustics or your local distributor for service if the problem
persists
Probe not detected. Check that the probe connector is fully seated in the
socket
Disconnect and reconnect the probe
Cycle instrument power
Contact Interacoustics or your local distributor for service if the problem
persists
Paired to Printer:
Check that the printer is turned on
o
Move closer to the printer
o
Try again
o
Paired to PC Computer or dongle:
Check that the serial port is open. Establish that the serial port is
o
handled by the PC and/or the software, not by the OtoRead
instrument
Check BT device (printer or PC) status
Attempt to connect to BT device again
Contact Interacoustics or your local distributor if the problem persists
Printing has been attempted, but no BT device is paired with the
TM
OtoRead
Instrument. Establish wireless pairing
The printer is not responding to queries from the instrument. Check
printer status
Awaken printer from sleep mode
Charge printer battery if necessary
DOWN arrow button for 10 seconds to force the
Page 45