Problem
Due for Service
Fit Error
Cannot Obtain L
Fit Error
Too High
Fit Error
Too Low
Limit
Error
Memory almost full
Memory Full!
Power Low!
Printer Error
Printer Paper Out!
Time/Date Error
Solutions
Indicates calibration of instrument is recommended. Message will
appear upon the calibration due date set in the device. Message
appears during device startup once per day
For a DP test, the desired level (L1 or L2) cannot be obtained
within allowable limits. User should refit the probe and retry the
test
Replace the probe tube
Contact Interacoustics or your local distributor for service if the
problem persists across several patients
For a DP test, the level of the calibration tone is too high. User
should refit the probe and retry the test
Replace the probe tube
Contact Interacoustics or your local distributor for service if the
problem persists across several patients
For a DP test, the level of the calibration tone is too low. User
should refit the probe and retry the test
Replace the probe tube
Contact Interacoustics or your local distributor for service if the
problem persists across several patients
Overflow error during the calculation of the DFTs for a DP test.
User should repeat the test
Cycle instrument power
Contact Interacoustics or your local distributor for service if the
problem persists
Saved tests are within 5 tests of the maximum limit. Print or
transfer test result to avoid interruption in testing
The maximum saved test limit is reached. The user will need to
clear the memory before any additional tests can be performed
The battery charge level is too low for operation. The user must
charge the battery before additional tests can be performed
Indicates a problem with the printer. Check the printer status.
Reset the printer or cycle the printer power
Replace the paper roll
The clock is checked during power on to ensure it has not lost
time and been reset. In the case of clock reset, this message is
shown. The user should set the correct date/time