Charging and powering the OtoRead

2.6.2.1 The battery

TM
The OtoRead
instrument is powered by an integrated rechargeable lithium-ion battery (1800 mAh)
providing 20 hours (500 tests, minimum) of operation between full charging.
The battery can be charged in two ways:
1. While placed in the powered cradle.
2. When connected via USB cable to the PC or to the provided mains adapter.
Please note that the battery life will vary depending on each product configurationand that the maximum
capacity of this battery will decrease with time and usage. To prolong the battery life, do not completely
deplete the battery, but charge when 5 %-10 % of the battery remains.
The probe houses a microphone and two speaker tubes which
produce test stimuli and measure the sound pressure level (SPL)
present in the sealed ear canal. Interface of the instrument to the ear
canal is accomplished through disposable ear tips made of industrial
elastomer which fit onto the probe tip. The disposable ear tips are
color coded to facilitate easy selection by size.
TM
The Micro-USB port on the bottom of the instrument is used for
charging via the power supply.
The blue battery indicator provides a visual indication (blue) of the
battery recharging function and battery status during operation.