3.11.5.8 Save protocol

Advanced options for TEOAE testing (TE menu)

3.11.6.1 General

Pleasse note: The Advanced Options for TEOAE Testing section is reserved for those units
purchased as a diagnostic device. Instruments purchased as a screener can skip this section.
The Advanced Options menu permits modification of the test stimuli and measurement values for the TE
Custom Protocol. Changes to the protocol should be made only by qualified personnel, usually the
administrator. If you are not familiar with the use of these variables, do not attempt to change the protocols.
Changes to any of these characteristics may yield test results that differ from those obtained in other test
modes.
TM
The OtoRead
instrument comes with pre-programmed protocol settings. See Appendix D for the
manufacturer settings of these protocols. Test protocol changes are saved in the non-volatile memory so the
settings will be retained even when changing the batteries.

3.11.6.2 Customizing a test protocol

To enter the TEOAE Menu:
1. Press
CHANGE at the main menu.
2. Using the CHANGE buttons, select the TEOAE protocol you want to customize (the 'TE 64s'
protocol is not customizable).
3. Press
SETUP at the Protocol menu.
4. At the Clock menu the teal arrow indicates that a custom protocol menu is available. Hold down
CHANGE key for 3 seconds until the 'READY' light (green LED) turns off.
5. At the New BT Device menu the teal arrow indicates that a custom protocol menu is available. Hold
down the
CHANGE key for 3 seconds until the 'READY' light (green LED) turns off.
You will now see the Averaging Time screen. You are in the TEOAE menu and will be able to scroll through
the available protocol parameters with the
arrow keys to CHANGE the selection.
If you push the
DOWN arrow key without holding it for 3 seconds, you will scroll through date and time,
etc., rather than accessing the custom protocol menus.
Once all of the settings have been selected for the protocol, these
settings can be saved by selecting the SAVE key. Press the
DONE key to exit.
Selection of
DONE before saving will not save the selection made
for the protocol.
The TEOAE settings can be recognized by the abbreviation 'TE' in the
top left corner.
NEXT button and make changes by using the LEFT or RIGHT