Battery Lifetime And Charging Time; Indication Lights On Otoread; Connecting The Micro-Probe To The Otoread Tm - Interacoustics OtoRead Instructions For Use Manual

Hide thumbs Also See for OtoRead:
Table of Contents

Advertisement

OtoRead™ Instructions for Use - US

2.6.2.2 Battery lifetime and charging time

CAUTION
When fully charged, the OtoRead
when the default 'power save' and 'power off' options are enabled.

2.6.2.3 Indication lights on OtoRead

2.6.2.4 Connecting the Micro-Probe to the OtoRead
Please observe the following precautions:
The battery can only be maintained by service personnel.
Keep the battery fully charged
Do not place the battery in a fire or apply heat to the battery
Do not damage the battery or use a damaged battery
Do not expose the battery to water
Do not short circuit the battery or reverse the polarity
The battery automatically starts charging when the OtoRead
placed properly into the AC-powered cradle or when connected
via USB to the PC
Use only the charger provided with the OtoRead
TM
The OtoRead
will shut down automatically when the battery charge is
too low to support continued operation.
TM
battery will last for a full day of testing under normal test conditions and
TM
During battery charging the indicator will be lit whenever the Micro-
USB connector is engaged and powered. The rate of illumination of
the indicator provides a means of identifying the status of the charging
function, and is defined as follows:
Steady-state illumination indicates that the battery is fully
charged. This identifies that the charging cycle has been
completed or was not implemented because the battery was
already fully charged
Slow blink illumination indicates that the charging function is
in process
Fast blink illumination indicates a fault condition in which the
user would be directed by the Operating Instructions to return
the instrument for service
During instrument operation, the user is warned of a low battery
condition by the following illumination of the CHARGE indicator: Two
fast blinks followed by a pause and then repeated until the battery is
charged.
TM
Turn off the OtoRead
socket on the top of the OtoRead
orientation.
An Interacoustics logo is on the probe connector and will align with the
instrument control panel.
Misalignment of the plug and socket can cause damage. The plug and
socket should be visually inspected prior to each installation of the remote
probe.
TM
and insert the Micro-Probe connector into the
TM
. The plug will fit in only one
Page 15
TM
is