The probe tube

Probe tubes are disposable and should be replaced when they become clogged. A package of replacement
probe tubes is included with this instrument. Do not attempt to clean the probe tube.
CAUTION
Menu structure
The OtoRead
TM
starts in the Main Menu. The following can be viewed or initiated from this screen:
The basic workflow is displayed in the flowchart underneath. From the Main Menu the connection to the
printer can be accessed by pressing the
selected test protocol, which can be seen in the Main Menu. Selecting the
the settings.
To replace the probe tube, use the ear tip to grasp the probe tube (the
clear plastic tube) and twist slightly while pulling the probe tube
straight out of the probe head.
Dispose of the used probe tube immediately to avoid confusing used
tubes with new tubes. Take a new probe tube from the package and
insert the tube into the probe head until it is fully seated. A properly
inserted probe tube will snap securely into place when it is fully seated
in the probe head.
If the probe tube is removed from the probe head and then re-used, it
will not fit as well as on its initial placement.
DO NOT ATTEMPT TO CLEAN PROBE TUBES. THIS MAY CAUSE
DAMAGE TO THE PROBE!
UP arrow. Selecting the
LEFT or RIGHT arrow will start the
SETUP arrow brings you to