7. Specifications
SpectraMax M2 & SpectraMax M2e Multi-mode Plate Readers User Guide — 0112-0102 Rev. D
48
Microplate read time (endpoint,
SCAN SPEED
Cuvette: Normal scan
Cuvette: Speed scan
Wavelength repeatability
TEMPERATURE REGULATION
Reading chamber
Temperature uniformity at equilibrium
Chamber warm-up time
Temperature regulation
Temperature regulation diagnostics
Recommended microplate
AUTOMIX WITH SOFTMAX PRO
• 96 wells in 18 seconds (single
Speed read
wavelength, absorbance)
• 384 wells in 49 seconds (single
wavelength, absorbance)
45*K nm/min (K = wavelength interval)
130*K nm/min
±0.2 nm
Isothermal when temperature regulation is
not enabled
Range
4°C above ambient to 45°C when temperature
regulation enabled. The ambient temperature
must be > 20°C to achieve temperature regu-
lation at 45°C.
Resolution
±0.1°C
Accuracy
±1.0°C for microplate and cuvette chamber
±0.5°C at 37°C
15–30 minutes (measured on air) after
initiation of temperature regulation
4 sensors
Drift
±0.2°C (regulated)
Temperature regulation system is continu-
ously monitored and updated
Evaporation
Plate lid required to minimize evaporative
cooling
Flat-bottom microplates with isolated wells
and lid
Control
Front panel reports cuvette chamber temper-
ature only (temperature for microplate cham-
ber reported in SoftMax Pro)