6. Troubleshooting
SpectraMax M2 & SpectraMax M2e Multi-mode Plate Readers User Guide — 0112-0102 Rev. D
44
Error Code
Error Message
500–599: NVRAM ERRORS
500
NVRAM CRC corrupt
501
NVRAM Grating cal data
bad
502
NVRAM Cuvette air cal
data error
503
NVRAM Plate air cal data
error
504
NVRAM Carriage offset
error
505
NVRAM Stage offset error
506
NVRAM Battery
For all other error messages (codes not listed here), please contact your local Molecular
Devices representative for assistance.
Notes
The CRC for the NVRAM data is corrupt.
Grating calibration data is unreasonable.
Cuvette air calibration data is unreasonable.
Plate air calibration data is unreasonable.
Carriage offset data is unreasonable.
Stage offset data is unreasonable.
Time to replace the NVRAM battery (U3).