Error Code
307
308
309
312
313
314
315
316
317
400–499: MOTION ERRORS
400
401
402
403
SpectraMax M2 & SpectraMax M2e Multi-mode Plate Readers User Guide — 0112-0102 Rev. D
Error Message
cuv air ref fail
stray light
front panel not respond-
ing
gain calibration failed
reference gain check fail
low lamp level warning
can't find zero order
grating motor driver
faulty
monitor ADC faulty
carriage motion error
filter wheel error
grating error
stage error
6.2. Error Codes and Probable Causes
Notes
Light leak in reading chamber or cuvette
door open. Could also be a faulty pre-amp
board.
LCD front panel bad or disconnected.
Power-up calibration and check of signal
path gain is out of tolerance. Could be due
to bad or disconnected pre-amp or
excessive stray light.
Power-up check of the Reference
amplifier's gain out of tolerance. Could be
due to bad or disconnected pre-amp board
or excessive stray light.
On power-up, grating motor could not find
zero-order home position.
Grating motor didn't move to where it was
commanded to in a reasonable time.
Carriage did not move to either of its
photo interrupts in a reasonable time, or
can't find its photo interrupt.
Filter wheel did not move to its photo
interrupt in a reasonable time, or can't find
its photo interrupt.
Grating did not move to its photo interrupt
in a reasonable time, or can't find its photo
interrupt.
Stage did not move to its photo interrupt
in a reasonable time, or can't find its photo
interrupt.