Supply room
VEGAPULS 64 • HART and accumulator pack
2
+
( )
Fig. 29: Supply room
1
Internal connection of the socket for the battery charger
2
Mode switch
3
LED green, charging process
4
LED yellow, charging status
5
Socket for the battery charger
The mode switch enables the selection of the following modes:
0 = sensor off, LEDs show the accumulator status
1, 2 = sensor off, LEDs off
3 = sensor on for 1 hour after pressing a key (delivery status)
4 = sensor permanently on, switching on/off via button
5 = sensor is switched on every 30 minutes for 3 minutes
6 = sensor is switched on every hour for 3 min.
7 = sensor is switched on every 6 hours for 3 minutes
8 = sensor is switched on every 12 hours for 3 minutes
9 = sensor is switched on every 24 hours for 3 minutes
The green LED characterizes the charging process:
LED flashes = Accumulator is charging
LED lights = accumulator is full, battery charger should be
unplugged (accumulator life time)
After pressing the key or changing the mode the yellow LED shows
the accumulator status for approximately 10 s as follows:
LED lights = accumulator is full
LED flashes = accumulator should be charged
LED off = accumulator is empty
5.3

Switch-on phase

The instrument is switched on and off by means of a button outside
on the housing.
5 Connecting to power supply
1,2 OFF
3 ON/OFF 1h
4 ON/OFF
0
5..9 auto cycle
Charge
modes
1
2
(-)
1
3
4
Accu
5
29