Time functions
Measured value indica-
tion
Selection of national
language
VEGAPULS 64 • HART and accumulator pack
5
4
Fig. 33: Display and adjustment elements - with adjustment via magnetic pen
1
LC display
2
Magnetic pen
3
Adjustment keys
4
Bluetooth symbol
5
Lid with inspection window
When the [+] and [->] keys are pressed quickly, the edited value,
or the cursor, changes one value or position at a time. If the key is
pressed longer than 1 s, the value or position changes continuously.
When the [OK] and [ESC] keys are pressed simultaneously for more
than 5 s, the display returns to the main menu. The menu language is
then switched over to "English".
Approx. 60 minutes after the last pressing of a key, an automatic reset
to measured value indication is triggered. Any values not confirmed
with [OK] will not be saved.
6.3
Measured value indication - Selection of
national language
With the [->] key you move between three different indication modes.
In the first view, the selected measured value is displayed in large
digits.
In the second view, the selected measured value and a correspond-
ing bar graph presentation are displayed.
In the third view, the selected measured value as well as a second se-
lectable value, e.g. the temperature of the electronics, are displayed.
During the initial setup of an instrument shipped with factory settings,
use the "OK" key to get to the menu "National language".
This menu item is used to select the national language for further pa-
rameter adjustment. You can change the selection via the menu item
"Setup - Display, Menu language".
6 Set up with the display and adjustment module
1
2
3
33

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents