Vega VEGAPULS 64 Operating Instructions Manual page 59

Hart and accumulator pack
Hide thumbs Also See for VEGAPULS 64:
Table of Contents

Advertisement

Check the 4 ... 20 mA
signal
Treatment of measure-
ment errors with liquids
VEGAPULS 64 • HART and accumulator pack
Connect a multimeter in the suitable measuring range according to
the wiring plan. The following table describes possible errors in the
current signal and helps to eliminate them:
Error
4 ... 20 mA signal not
stable
4 ... 20 mA signal miss-
ing
Current signal greater
than 22 mA, less than
3.6 mA
The below tables show typical examples of application-related meas-
urement errors with liquids. The measurement errors are differentiated
according to the following:
Constant level
Filling
Emptying
The images in column "Error pattern" show the real level as a broken
line and the level displayed by the sensor as a continuous line.
0
1
Real level
2
Level displayed by the sensor
Notes:
Wherever the sensor displays a constant value, the reason could
also be the fault setting of the current output to "Hold value"
If the level indication is too low, the reason could be a line resist-
ance that is too high
9 Diagnosis, asset management and service
Cause
Fluctuating measured
value
Electrical connection
faulty
Voltage supply missing
Operating voltage too
low, load resistance
too high
Sensor electronics
defective
Rectification
Set damping
Check connection, cor-
rect, if necessary
Check cables for
breaks; repair if neces-
sary
Check, adapt if neces-
sary
Exchange the instru-
ment or send it in for
repair
1
2
time
59