Advertisement

Code
Cause
Text message
F125
Impermissible elec-
tronics temperature
F260
Error in the cali-
bration
F261
Error in the instru-
ment settings
F264
Installation/Setup
error
F265
Measurement func-
tion disturbed
Function check
Code
Cause
Text message
C700
Simulation active
Out of specification
Code
Cause
Text message
S600
Impermissible elec-
tronics temperature
S601
Overfilling
S603
Impermissible oper-
ating voltage
VEGAPULS 64 • HART and accumulator pack
Temperature of the electronics
in the non-specified range
Error in the calibration carried
out in the factory
Error in the EEPROM
Error during setup
False signal suppression faulty
Error when carrying out a reset
Adjustment not within the vessel
height/measuring range
Max. measuring range of the
instrument not sufficient
Sensor no longer carries out a
measurement
Operating voltage too low
The following table shows the error codes and text messages in the
status message "Function check" and provides information on causes
as well as corrective measures.
A simulation is active
The following table shows the error codes and text messages in the
status message "Out of specification" and provides information on
causes as well as corrective measures.
Temperature of the electronics
in the non-specified range
Danger of vessel overfilling
Operating voltage below speci-
fied range
9 Diagnosis, asset management and service
Rectification
Check ambient temperature
Insulate electronics
Use instrument with higher
temperature range
Exchanging the electronics
Send instrument for repair
Repeat setup
Carry out a reset
Check or correct installation
and/or parameter settings
Use an instrument with bigger
measuring range
Check operating voltage
Carry out a reset
Disconnect operating voltage
briefly
Rectification
Finish simulation
Wait for the automatic end after
60 mins.
Rectification
Check ambient temperature
Insulate electronics
Use instrument with higher
temperature range
Make sure that there is no
further filling
Check level in the vessel
Check electrical connection
If necessary, increase operating
voltage
DevSpec
State in CMD 48
Bit 7 of Byte 0 ... 5
Bit 8 of Byte 0 ... 5
Bit 9 of Byte 0 ... 5
Bit 10 of Byte 0 ... 5
Bit 11 of Byte 0 ... 5
DevSpec
State in CMD 48
"Simulation Active"
in "Standardized
Status 0"
DevSpec
State in CMD 48
Bit 8 of
Byte 14 ... 24
Bit 9 of
Byte 14 ... 24
Bit 11 of
Byte 14 ... 24