6. Physical installation stand alone cooler
Details and limitations
Depending the mounting option chosen, some limitations must be considered:
Floor mounting: When this option is chosen, the cooler is installed on a flat and stable surface, in horizontal position. Four
feet correctly distributed (2 on each side) allow stabilize the device. The cooler could be tilted forward or backward, but for
security reasons it is strongly recommended to let the unit in horizontal position, the best way to avoid the potential hazards of
falls and slips.
Image 6-2
1
Four feet allow stabilize the unit.
Stacking Standalone coolers: It is NOT allowed to stack multiple cooler modules on top of each other. The air intake of the
upper cooler would take in the "hot" air of the lower cooler, even at a distance of 1m apart.
Image 6-3
A Stack of cooler positioned on their sides may be considered, but in this case, the direction of the entire coolers need to be
the same, in order to forcing air flow in the same direction.
Image 6-4
Note: This positioning must be secularized with an adapted structure.
52
360°
x4
1
R5906753 DPXK BLP SERIES 20/11/2017