11.1 Output / Output rating / Hardware adjustments

RS-232 and USB:
- 8 data bits
- 1 Start bit, 1 Stop bit
- No parity
- Baud rate: 9600
- Peak Capture Rate: 0.100 S
- ADC Sampling Rate: 1,000 Hz
11.2 RS 232 interface occupancy
12 Adjustment procedure
Introduction
This instruction describes the calibration procedure, where a calibration password
has to be entered, to enter certain modes of the software.
Note: Throughout this procedure, all masses used to apply a load to the load cell,
should be calibrated.
Temperature Acclimatization
Inspect the instrument for damages and check all ordered requirements before
placing in the calibration laboratory. Before any calibration can start, the instrument
has to become acclimatized to the operating temperature of the calibration laboratory
for at least two hours.
Preparation
Fit a full set of charged batteries into the gauge's battery compartment or plug in a
standard power supply.
FL-BA-e-1714
13