page. If you selected DELETE ALL and press ENTER key the gauge will delete all
saved record and return to memory page.
4) ADJUSTMENT
Experienced staff should adjust the instrument only.
To get into the adjustment mode, you need to enter a password, which is is 7780.
You can find a detailed description of this procedure in chapter 12 at the end of this
instruction manual. Take care that you need weights to adjust the force gauge. For
more details, please contact your SAUTER distributor or SAUTER GmbH directly per
phone (number mentioned above) or by e-mail.
5) DIAGNOSTIC
This is used to check the status of the load cell. If you suspect that your load cell
transducer has sustained an overload, it is possible to check the status of the load
cell immediately.
Place the gauge horizontally on the flat level surface and go to main menu page. Use
UP and DOWN key to move the cursor point to DIAGNOSTIC and press ENTER key
the display will show Diagnostic menu page. Press ESC to return to main menu
page.
Figure 8 Diagnostic Menu
If the % offset is between 5% - 10 % please contact your supplier to arrange a
recalibration of your gauge.
If the % offset is greater than 10% please contact your supplier to arrange for load
cell replacement.
These values are given as an indicator only – the need for calibration / repair may
vary according to the individual characteristics of the load cell.
6) SELECT OUTPUT
In this menu, you can choose the output port between USB or RS232
FL-BA-e-1714