Press UP and DOWN to change max capacity, press and hold to scroll change.
Press ENTER key to save selected and return to Calibration Menu page.
4) Tension Calibration. At the Calibration menu page, press UP and DOWN key to
move the cursor to point to TENSION GAIN and press ENTER key. Tension gain
menu page will be displayed:
Fig. 5 Tension Gain Menu page
4.1) Tension Zero
- Set the gauge up with the necessary supports to carry out tension calibration
- Pre-stress the load cell to 100% of the instrument's capacity.
- Remove the masses but retain the supports.
- When the counts have settled, press ZERO key to store the zero counts
4.2) Tension Maximum
- Apply 100% of the instrument's capacity load onto the support(s) for tension.
- Press UP, DOWN, RIGHT ARROW or LEFT ARROW key to adjust tension gain
until the gauge displayed max capacity count.
- Press ENTER key to store the counts for full-scale tension.
5.) Compression Calibration. At Calibration menu page, press UP and DOWN key to
move the cursor point to COMPRESSION GAIN and press ENTER key. The display
will show compression gain menu page.
Fig. 6 Compression Gain Menu page
5.1) Compression Zero
- Set the gauge up with the necessary supports to carry out compression calibration.
- Pre-stress the load cell to 100% of the instrument's capacity.
- Remove the masses but retain the supports.
- When the counts have settled, press ZERO key to store the zero counts
5.2) Compression Maximum
FL-BA-e-1714
15