Output Protocol - sauter FH-M 1 kN Instruction Manual

Fh-m series. digital force gauge
Table of Contents

Advertisement

Pin
Signal
Illustration
2
TxD
Output signal
3
RxD
Input signal
5
GND
Ground
6
+1.6 to +
>upper limit value
2 V
7
+1.6 to +
< lower limit value
2 V
8
+1.6 to +
OK
2 V

7.1 Output Protocol

RS-232 Parameter
-
Baudrate:
-
Data-Bit:
-
Parity:
-
Stop-bit:
The measured value is requested by the PC by the ASCII Sign "9".
The measured value that comes from the instrument has this format:
e.g.
0011.70
||-----|
|
|____>
Signs
|_______>
minus = Pressure;
1 = plus = Tension)
or:
1021.15 means +21,15 N (in tension direction)
|_______>
Pressure;
1 = plus = Tension)
or:
1021.15 means +21,15 N (in tension direction)
FH-M-BA-e-1718
9600
8
none
1
means -11.70 Newton, if Newton is the selected unit
the other 6 places define the measured value as ASCII-
the first place describes the direction of the force (0 =
the first place describes the direction of the force (0 = minus =
9

Advertisement

Table of Contents
loading

This manual is also suitable for:

Fh-m2 knFh-m5 knFh-m10 knFh-m20 knFh-m50 knFh-m 100 kn