3. Hearing Instrument Style:
Select the type of hearing aid the client is fitted with. Choose between Behind the ear, In the
canal, In the ear, and Completely in the canal. This will add a correction to the measurement
during RECD based verification in the test box.
4. Hearing Aid Type:
Choose the type of amplification. Select between Linear and Nonlinear using the check boxes.
5. Tubing:
Select the hearing aid tubing. Choose from Libby 4, Libby 3, CFA#2 Horn, CFA#3 stepped bore,
#13,6CS, 6C10, None.
6. Limiting:
Set the limiting to Off, Wide-band, and Multi-band. Multi-band will automatically be selected as
standard.
Note that limiting information regarding individual hearing aids may be found in the
manufacturers' specification sheets or by contacting the manufacturer.
7. Orientation for REUR/REUG:
If using a predicted REUR/REUG select orientation. Choose between 0 Deg and 45 Deg using
the check boxes. 90 Deg is not available for NAL-NL2
8. Vent Size:
Set the vent size of the hearing aid using the dropdown. Select between Tight, Occluded, 1 mm,
2 mm, 3 mm, and Open.
9. Number of Hearing Aids:
Set whether the fitting is Monaural or Binaural using the tick boxes
10. Threshold (dB)
Type in the preferred knee-point in the white section which is set to 50 by default
Audiogram type
Select to use bone conduction values in the target calculation.

2.6.4.4 Print Layout Settings

The REM440 module allows for setting up individual print layouts. This is done in the Print Wizard which is
entered using the Print Wizard button (read about the Print Wizard in section 4.1).