Reig Settings - Interacoustics Affinity 2.0 Additional Information

Analyzer
Hide thumbs Also See for Affinity 2.0:
Table of Contents

Advertisement

2.0
Affinity
Additional Information

2.6.2.3 REIG Settings

1
1) Ensure that REIG is marked in the left hand side of the screen. The name of the test will then appear
in the Name of selected test section.
Note the option to write a new name for the test in this section (e.g. "REIG 70 dB input")
2) Stimulus Selection:
Input Level: Set the input level using the slider.
Smoothing Index: You have the option to smooth the measured curve using the slider (see
section 0 for details). The higher index number selected the more smoothing is applied.
Continuous: This check box allows you to make the measurement ongoing rather than a single
sweep. This may, however, also be done in the front screen using the continuous button:
When continuous is ticked, the stimulus and recording will continue until you press the STOP
button. This can be practical either if you want to change the settings of the hearing aid or
observe the changes in hearing aid performance.
View as gain: allows you to view the measurement as a gain curve by default. If you wish to
view it as response, ensure that the check box remains unticked (see section 0 for details
regarding Gain versus Response view).
Measurement time (in seconds): Duration for the measurement can be set using the slider. This
is only relevant for broad band signals. A number of seconds can freely be selected between 1
and 1200. By default the measurement time is the same for all signals.
Instructions: This allows for adding picture instructions. If the box is ticked an instruction box
explaining how to perform the measurement will appear upon pressing START.