Page 164
Note that your choice has an impact on your result as illustrated below:
Warble
ICRA
Pink
IST
Note: The signals Male Speech, Female Speech, and ISTS are shaped according to LTASS standards.
Vocal efforts are included which results in the input level range being limited depending on the start level. If
the measurement is set to start at for example a normal input level (e.g. 65 dB) the input levels slider can
only be moved within normal vocal effort limits while measuring (40 – 70 dB). For a raised input level the
slider can be moved from 71 – 74 dB. If the measurement is set to start at a loud input level the slider can
only be moved within 75 – 82 dB and if the measurement is set to a shout level the input level slider can be
moved from 83 dB and above.

2.6.6.5 Presenting a Stimulus

The REM440 stimuli can be selected from the main screen using the stimuli selection dropdown
For continuous stimulus presentation, click the continuous button. When presenting continuously the curve
may be frozen in a certain position using the freeze curve button: