9) Vent Size:
Set the vent size of the hearing aid using the dropdown. Select between Tight, Occluded, 1 mm,
2 mm, 3 mm, and Open.
10) Number of Hearing Aids:
Set whether the fitting is Monaural or Binaural using the tick boxes
11) Threshold (dB)
Type in the preferred knee-point in the white section which is set to 50 by default
Audiogram type
Select to use bone conduction values in the target calculation.
12) Language
Select the type of language: Tonal or Non-Tonal. This will add a correction to the target
calculation based on whether or not a language is tonal (like many asian languages) or non-
tonal (like many European languages).

2.6.4.3 NAL-NL1 Settings

1. Default Prescription:
NAL-NL1.
2. Transducer:
Choose the type of transducer used for the audiometry. Select between Head phone, Insert,
Insert + mould, Sound field 0 deg., and sound field 45 deg. This will affect the threshold
indications and thereby the target calculation.
Note: It is important to choose the correct settings in this dropdown. E.g. if your audiogram is
done in head phones but inserts are selected as transducer, no target will be calculated.
Page 235