SE-1515 PC ECG User Manual
service department or your local distributor.
2. The system should be installed by a qualified service engineer. Do not power on the
system until all cables are properly connected and verified.
3. If multiple instruments are connected to a patient, the sum of the leakage currents may
exceed the limits given in the IEC/EN 60601-1 and may pose a safety hazard. Consult
your service personnel.
4. DX12 transmitter of the wireless system uses the Bluetooth technology, which could
make the patient with the pacemaker uncomfortable. Keep DX12 transmitter far away
from the pacemaker when using the wireless system of SE-1515.

2.2.2 Assembling Wireless System

DX12 Transmitter
Patient Cable
DX12 Receiver