SE-1515 PC ECG User Manual
When you click Re-identify after you have manually adjusted the measurement line, the system
automatically calculates the place of the measurement line of the Q point and J point and their
corresponding ST value and ST slope. The system also refreshes the place of the measurement
line of the Q point and J point every 10s.
1. single amplified mean wave
You can manage the lead to be displayed by system scan or manual configuration.
2. multiple original mean waves
In the manual mode, double-click on any mean wave area and the system will automatically
switch to its enlarged display window.
3. ST baseline
Right-click on the area of a single amplified mean wave or multiple original mean waves and
the system will display the Accumulate Baseline/Accumulate All.
Select the Accumulate Baseline and the baseline mean wave will be folded up to the
average template of each lead. This operation is only available during the Exercise stage.
Select the Accumulate All and the mean wave of all leads will be folded up on the
single amplified mean wave window.
4.4.3.4 ST Trend
On the ST Trend screen, the ST run charts of 6 leads are displayed. You can right-click to switch
between lead groups.
4.4.3.5 Sample Setup
Click on Settings on the exercise ECG sampling screen, and the System Setting window will be
displayed. In the System Setting window, click Sample Setting to enter the Sample Setting
window.
In the Sampling Setting window, you can
Configure the model, port number, device ID of the sampling box.
Configure the parameters of the filter.
Configure the lead mode, lead sequence, lead electrode, display sequence and QRS voice.
In addition, the lead sequence can be edited.
Select the Enter the analysis screen when sampling finishes.
Select the Auto print when sampling finishes.
- 51 -
ECG Sampling