 On the magnified waveform screen, drag the mouse and an electronic measurement ruler
(hereinafter called ruler) and corresponding measurement data will be displayed. You
can move the ruler by pressing Up/Down/Left/Right arrow keys.
NOTE: The mark line and ruler cannot be used at the same time, please right
-click to disable either of them first. To enable it, right-click again.
8. VCG:
If Enable Vector Calculation is selected in the Function/Algorithm Settings window, you
can click VCG for vector calculation on the resting ECG analysis screen. For details, refer to
section 5.3 "VCG".

5.1.2 Average Template

Click Template to enter the Average Template window for resting ECG. In this window, you can
analyze the data of waveforms on the Average Template.
 When you press ALL, magnified average templates of all leads will be overlapped with the
same central axis.
 You can set the speed and the gain of average templates.
 You can drag marker lines of P1, P2, Q, S and T on average templates.
P1 is the onset of P wave, P2 is the offset of P wave, Q marks the onset of the QRS wave, S
marks the offset of the QRS wave, and T is the offset of the T wave. You can move these
lines by dragging on the mouse and the corresponding parameter values will change. You can
also use the arrows key on the keyboard to move these marks.
 Reset
After you manually modify the position of a marker line, you can click Reset to restore it to
the initial position. The ST and ST slope will be automatically updated at the same time.
 Re-diagnose
After you manually modify the position of a marker line, the diagnosis result which is based
on the marker line position and corresponding measurement information will not update
automatically, you have to click Re-diagnose to update to update the diagnosis result.

5.1.3 About the Detail Information Window

On the Average Template window for resting ECG, click on Detail Information at the bottom
left corner to enter the Detail Information screen.
You can click Export Excel to export an Excel file.