Parameter
Description
Enable Vector
Once enabled, you can view the vector calculation results on the analysis
Calculation
screen for resting ECG.
NOTE: This function
Parameter
Setting
ECG key
Options provided are Start and Forbidden.
When Start is selected, the ECG key on the DE15/DE18 sampling box
functions as the Start button on the ECG sampling Screen.
When Forbidden is selected, there will be no response when pressing the
ECG key.
4.4.1.7 Other Setup
Click on Setting on the resting ECG sampling screen, and the System Setting window will be
displayed. In the System Setting window, click Others to enter the Others screen.
Item
Description
Display
Standard+XYZ mode..
Axis Calculation Method can be set to Area Method or Amplitude
Method.
Tachycardia Criterion (greater than): Manual input, default: 100
bpm.
When the patient's heart rate exceeds the Tachycardia Criterion, a
Tachycardia hint will appear in the diagnosis result, and the HR
information will be stressed with red.
Bradycardia Criterion (less than) : Manual input, default: 60 bpm
When Anti-aliasing is selected, the waveforms displayed or printed will
be smoother.
When 1mV marker is enabled, the 1mV calibration mark will be
displayed at the start of a line of waveforms in the sampling screen or
analysis screen.
Grid View can be set to:
5mm: Only the 5mm grid is dispplayed on the waveform screens.
The 1mm grid will not be displayed.
1mm: Both 5mm grid and 1mm grid will be displayed on the
waveform screens.
1s/1mV: On the waveform screens, 1s is regarded as a grid (5mm)
horizontally, and 1mV is regarded as a grid (5mm) vertically.
is unavailable in
- 46 -
ECG Sampling
9-Lead mode and