EDAN SE-1515 User Manual page 107

Diagnostic ecg
Table of Contents

Advertisement

SE-1515 PC ECG User Manual
Electrical
fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Power frequency
(50Hz/60Hz)
magnetic field
IEC 61000-4-8
Voltage dips, short
interruptions
and
voltage variations on
power supply input
lines
IEC 61000-4-11
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
Guidance and manufacture's declaration - electromagnetic immunity
SE-1515 PC ECG is intended for use in the electromagnetic environment specified below. The
customer or the user of SE-1515 PC ECG should assure that it is used in such an environment.
2
kV for
power
supply lines
1 kV line to line
2 kV line to ground
3A/m
<5% U
T
(>95% dip in U
)
T
for 0.5 cycle
40% U
T
(60% dip in U
)
T
for 5 cycles
70% U
T
(30% dip in U
)
T
for 25 cycles
<5% U
T
(>95% dip in U
)
T
for 5 sec
Electromagnetic Immunity
Not applicable
Not applicable
3A/m
Not applicable
- 100 -
EMC Information
Not applicable
Not applicable
Power frequency magnetic
fields should be at levels
characteristic of a typical
location
in
a
commercial
or
hospital
environment.
Not applicable