SE-1515 PC ECG User Manual
4.4.3.7 Protocol Editing
In the Device window, select a protocol and click Edit to enter the editing window for
non-default protocols. In this window, you can change any parameter of each stage or delete the
non-default protocols.
4.4.3.8 Parameter Setup
Item
Target HR
Normal BP range
POST J
ST Standards
Manual Report
Auto Report
Description
You can configure the Max predicted HR and the calculation method of
the target HR.
You can configure the systolic and diastolic to the value in the normal
range.
If the blood pressure of the patient excesses the configured value, an
indication message will be displayed on the main screen and the
background color of the blood pressure parameter will be yellow.
It can be set to Manual or Auto.
NOTE: J point is the endpoint of the QRS waveform group and
the startpoint of the ST segment. It is also the reference
point used by the system to determine the place of the
ST segment. Please select the J point based on the
actual ECG waveforms of the patient.
When ST Overrun hint is selected, the system will display an
indication message ST Overrun when the ST value is detected to be
beyond the threshold.
You can set the threshold for ST Elevation or ST Depression only
after STOverrun hint is selected.
ST Elevation can be set to 0.05 ~0.3mV, and ST Depression can be set
to -0.05 ~-0.3mV.
Options provided are Print, Print and Save, Save.
When Print/Print and Save/Save is selected, you can click Print to
Print/Print and Save/Save the corresponding ECG report during the
exercise test.
Options provided are OFF, Print, Print and Save, Save.
If OFF is selected, the 12-lead report will not be printed automatically
on the scheduled auto-print time (configured on the Edit Protocol
window) during the exercise test.