SE-1515 PC ECG User Manual
Appendix 2 EMC Information
Guidance and manufacture's declaration - electromagnetic emission
SE-1515 PC ECG is intended for use in the electromagnetic environment specified below. The
customer or the user of SE-1515 PC ECG should assure that it is used in such an environment.
Emission test
RF emissions
CISPR 11
RF emission
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage
fluctuations/
flicker emissions
IEC 61000-3-3
Guidance and manufacture's declaration - electromagnetic immunity
SE-1515 PC ECG is intended for use in the electromagnetic environment specified below. The
customer or the user of SE-1515 PC ECG should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electromagnetic Emissions
Compliance
Electromagnetic environment - guidance
SE-1515 PC ECG uses RF energy only for its internal
function. Therefore, its RF emissions are very low and
Group 1
are not likely to cause any interference in nearby
electronic equipment.
Class A
SE-1515 PC ECG is suitable for use in all
establishments, other than domestic and those directly
Not applicable
connected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes.
Not applicable
Electromagnetic Immunity
IEC 60601 test level
6 kV contact
8 kV air
Compliance level
6 kV contact
8 kV air
- 99 -
EMC Information
Electromagnetic
environment - guidance
Floors should be wood,
concrete or ceramic tile. If
floor
are
covered
synthetic
material,
relative humidity should be
at least 30%.